User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Tektronix 2232 Sub7 DATE: 12-Aug-93 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 49 NUMBER OF LINES: 347 CONFIGURATION: Fluke 6060B ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 ASK+ X #----------------------------- 60MHz ----------------------------------------- 1.003 HEAD {} 1.004 HEAD {Can a stable 60MHz 1 division display be obtained} 1.005 HEAD {adjusting the A TRIGGER LEVEL control?} 1.006 HEAD {} 1.007 HEAD {CH1, A trigger mode P-P AUTO, slope positive} # This jmp FSC is needed for Cancel and Repeat options during test run. 1.008 JMP 2.001 1.009 EVAL Dummy 2.001 HEAD Internal A and B Triggering Test 2.002 DISP Move the Fluke 6061A RF OUTPUT via a 50ê feedthru 2.002 DISP termination to the UUT CH1 input connector. 2.003 DISP Set the UUT VERTICAL controls as follows: 2.003 DISP 2.003 DISP ÿ MODE ........................................... CH1 2.004 DISP Set the UUT HORIZONTAL controls as follows: 2.004 DISP 2.004 DISP ÿ MODE ............................................. A 2.004 DISP ÿ A & B SEC/DIV ................. (A & B LOCKED) 0.1æs 2.004 DISP ÿ X10 Magnifier ........................ On (knob out) 2.005 DISP Set the UUT B TRIGGER controls as follows: 2.005 DISP 2.005 DISP ÿ SLOPE ............................... + (button out) 2.006 6060 60MH 50mVpp -D S 2.007 MESS Can a stable display be obtained by adjusting the 2.008 EVAL A TRIGGER LEVEL control? 3.001 HEAD {} 3.002 HEAD {CH1, A trigger mode P-P AUTO, slope negative} # This jmp FSC is needed for Cancel and Repeat options during test run. 3.003 JMP 4.001 3.004 EVAL Dummy 4.001 HEAD Internal A and B Triggering Test 4.002 6060 60MH 50mVpp -D S 4.003 DISP Set the UUT A TRIGGER controls as follows: 4.003 DISP 4.003 DISP ÿ SLOPE ................................ - (button in) 4.004 EVAL A TRIGGER LEVEL control? 5.001 HEAD {} 5.002 HEAD {CH1, A trigger mode NORM, slope negative} # This jmp FSC is needed for Cancel and Repeat options during test run. 5.003 JMP 6.001 5.004 EVAL Dummy 6.001 HEAD Internal A and B Triggering Test 6.002 6060 60MH 50mVpp -D S 6.003 DISP Set the UUT A TRIGGER controls as follows: 6.003 DISP 6.003 DISP ÿ A TRIGGER MODE ................................ NORM 6.004 EVAL A TRIGGER LEVEL control? 7.001 HEAD {} 7.002 HEAD {CH1, A trigger mode NORM, slope positive} # This jmp FSC is needed for Cancel and Repeat options during test run. 7.003 JMP 8.001 7.004 EVAL Dummy 8.001 HEAD Internal A and B Triggering Test 8.002 6060 60MH 50mVpp -D S 8.003 DISP Set the UUT A TRIGGER controls as follows: 8.003 DISP 8.003 DISP ÿ SLOPE ............................... + (button out) 8.004 EVAL A TRIGGER LEVEL control? 9.001 HEAD {} 9.002 HEAD {CH1, B trigger mode NORM, slope positive} # This jmp FSC is needed for Cancel and Repeat options during test run. 9.003 JMP 10.001 9.004 EVAL Dummy 10.001 HEAD Internal A and B Triggering Test 10.002 6060 60MH 50mVpp -D S 10.003 DISP Set the UUT HORIZONTAL controls as follows: 10.003 DISP 10.003 DISP ÿ MODE ............................................. B 10.004 DISP Set the UUT A TRIGGER controls as follows: 10.004 DISP 10.004 DISP ÿ A TRIGGER MODE ............................ P-P AUTO 10.005 DISP Set the UUT B TRIGGER controls as follows: 10.005 DISP 10.005 DISP ÿ LEVEL ..................................... Midrange 10.006 EVAL B TRIGGER LEVEL control? 11.001 HEAD {} 11.002 HEAD {CH1, B trigger mode NORM, slope negative} # This jmp FSC is needed for Cancel and Repeat options during test run. 11.003 JMP 12.001 11.004 EVAL Dummy 12.001 HEAD Internal A and B Triggering Test 12.002 6060 60MH 50mVpp -D S 12.003 DISP Set the UUT B TRIGGER controls as follows: 12.003 DISP 12.003 DISP ÿ SLOPE ................................ - (button in) 12.004 EVAL B TRIGGER LEVEL control? 13.001 HEAD {} 13.002 HEAD {CH2, A trigger mode P-P AUTO, slope positive} # This jmp FSC is needed for Cancel and Repeat options during test run. 13.003 JMP 14.001 13.004 EVAL Dummy 14.001 HEAD Internal A and B Triggering Test 14.002 DISP Move the Fluke 6061A RF OUTPUT via a 50ê feedthru 14.002 DISP termination to the UUT CH2 input connector. 14.003 DISP Set the UUT VERTICAL controls as follows: 14.003 DISP 14.003 DISP ÿ MODE ........................................... CH2 14.004 DISP Set the UUT HORIZONTAL controls as follows: 14.004 DISP 14.004 DISP ÿ MODE ............................................. A 14.005 DISP Set the UUT B TRIGGER controls as follows: 14.005 DISP 14.005 DISP ÿ SLOPE ............................... + (button out) 14.006 6060 60MH 50mVpp -D S 14.007 MESS Can a stable display be obtained by adjusting the 14.008 EVAL A TRIGGER LEVEL control? 15.001 HEAD {} 15.002 HEAD {CH2, A trigger mode P-P AUTO, slope negative} # This jmp FSC is needed for Cancel and Repeat options during test run. 15.003 JMP 16.001 15.004 EVAL Dummy 16.001 HEAD Internal A and B Triggering Test 16.002 6060 60MH 50mVpp -D S 16.003 DISP Set the UUT A TRIGGER controls as follows: 16.003 DISP 16.003 DISP ÿ SLOPE ................................ - (button in) 16.004 EVAL A TRIGGER LEVEL control? 17.001 HEAD {} 17.002 HEAD {CH2, A trigger mode NORM, slope negative} # This jmp FSC is needed for Cancel and Repeat options during test run. 17.003 JMP 18.001 17.004 EVAL Dummy 18.001 HEAD Internal A and B Triggering Test 18.002 6060 60MH 50mVpp -D S 18.003 DISP Set the UUT A TRIGGER controls as follows: 18.003 DISP 18.003 DISP ÿ A TRIGGER MODE ................................ NORM 18.004 EVAL A TRIGGER LEVEL control? 19.001 HEAD {} 19.002 HEAD {CH2, A trigger mode NORM, slope positive} # This jmp FSC is needed for Cancel and Repeat options during test run. 19.003 JMP 20.001 19.004 EVAL Dummy 20.001 HEAD Internal A and B Triggering Test 20.002 6060 60MH 50mVpp -D S 20.003 DISP Set the UUT A TRIGGER controls as follows: 20.003 DISP 20.003 DISP ÿ SLOPE ............................... + (button out) 20.004 EVAL A TRIGGER LEVEL control? 21.001 HEAD {} 21.002 HEAD {CH2, B trigger mode NORM, slope positive} # This jmp FSC is needed for Cancel and Repeat options during test run. 21.003 JMP 22.001 21.004 EVAL Dummy 22.001 HEAD Internal A and B Triggering Test 22.002 6060 60MH 50mVpp -D S 22.003 DISP Set the UUT HORIZONTAL controls as follows: 22.003 DISP 22.003 DISP ÿ MODE ............................................. B 22.004 DISP Set the UUT A TRIGGER controls as follows: 22.004 DISP 22.004 DISP ÿ A TRIGGER MODE ............................ P-P AUTO 22.005 EVAL B TRIGGER LEVEL control? 23.001 HEAD {} 23.002 HEAD {CH2, B trigger mode NORM, slope negative} # This jmp FSC is needed for Cancel and Repeat options during test run. 23.003 JMP 24.001 23.004 EVAL Dummy 24.001 HEAD Internal A and B Triggering Test 24.002 6060 60MH 50mVpp -D S 24.003 DISP Set the UUT B TRIGGER controls as follows: 24.003 DISP 24.003 DISP ÿ SLOPE ................................ - (button in) 24.004 EVAL B TRIGGER LEVEL control? #------------------------------- 100MHz -------------------------------------- 25.001 HEAD {} 25.002 HEAD {Can a stable 100MHz 1.5 division display be obtained} 25.003 HEAD {adjusting the TRIGGER LEVEL control?} 25.004 HEAD {} 25.005 HEAD {CH1, A trigger mode P-P AUTO, slope positive} # This jmp FSC is needed for Cancel and Repeat options during test run. 25.006 JMP 26.001 25.007 EVAL Dummy 26.001 HEAD Internal A and B Triggering Test 26.002 DISP Move the Fluke 6061A RF OUTPUT via a 50ê feedthru 26.002 DISP termination to the UUT CH1 input connector. 26.003 DISP Set the UUT VERTICAL controls as follows: 26.003 DISP 26.003 DISP ÿ MODE ........................................... CH1 26.004 DISP Set the UUT HORIZONTAL controls as follows: 26.004 DISP 26.004 DISP ÿ MODE ............................................. A 26.004 DISP ÿ A & B SEC/DIV ................ (A & B LOCKED) 0.05æs 26.005 DISP Set the UUT B TRIGGER controls as follows: 26.005 DISP 26.005 DISP ÿ SLOPE ............................... + (button out) 26.006 6060 100MH 100mVpp -D S 26.007 MESS Can a stable display be obtained by adjusting the 26.008 EVAL A TRIGGER LEVEL control? 27.001 HEAD {} 27.002 HEAD {CH1, A trigger mode P-P AUTO, slope negative} # This jmp FSC is needed for Cancel and Repeat options during test run. 27.003 JMP 28.001 27.004 EVAL Dummy 28.001 HEAD Internal A and B Triggering Test 28.002 6060 100MH 100mVpp -D S 28.003 DISP Set the UUT A TRIGGER controls as follows: 28.003 DISP 28.003 DISP ÿ SLOPE ................................ - (button in) 28.004 EVAL A TRIGGER LEVEL control? 29.001 HEAD {} 29.002 HEAD {CH1, A trigger mode NORM, slope negative} # This jmp FSC is needed for Cancel and Repeat options during test run. 29.003 JMP 30.001 29.004 EVAL Dummy 30.001 HEAD Internal A and B Triggering Test 30.002 6060 100MH 100mVpp -D S 30.003 DISP Set the UUT A TRIGGER controls as follows: 30.003 DISP 30.003 DISP ÿ A TRIGGER MODE ................................ NORM 30.004 EVAL A TRIGGER LEVEL control? 31.001 HEAD {} 31.002 HEAD {CH1, A trigger mode NORM, slope positive} # This jmp FSC is needed for Cancel and Repeat options during test run. 31.003 JMP 32.001 31.004 EVAL Dummy 32.001 HEAD Internal A and B Triggering Test 32.002 6060 100MH 100mVpp -D S 32.003 DISP Set the UUT A TRIGGER controls as follows: 32.003 DISP 32.003 DISP ÿ SLOPE ............................... + (button out) 32.004 EVAL A TRIGGER LEVEL control? 33.001 HEAD {} 33.002 HEAD {CH1, B trigger mode NORM, slope positive} # This jmp FSC is needed for Cancel and Repeat options during test run. 33.003 JMP 34.001 33.004 EVAL Dummy 34.001 HEAD Internal A and B Triggering Test 34.002 6060 100MH 100mVpp -D S 34.003 DISP Set the UUT HORIZONTAL controls as follows: 34.003 DISP 34.003 DISP ÿ MODE ............................................. B 34.004 DISP Set the UUT A TRIGGER controls as follows: 34.004 DISP 34.004 DISP ÿ A TRIGGER MODE ............................ P-P AUTO 34.005 DISP Set the UUT B TRIGGER controls as follows: 34.005 DISP 34.005 DISP ÿ LEVEL ..................................... Midrange 34.006 EVAL B TRIGGER LEVEL control? 35.001 HEAD {} 35.002 HEAD {CH1, B trigger mode NORM, slope negative} # This jmp FSC is needed for Cancel and Repeat options during test run. 35.003 JMP 36.001 35.004 EVAL Dummy 36.001 HEAD Internal A and B Triggering Test 36.002 6060 100MH 100mVpp -D S 36.003 DISP Set the UUT B TRIGGER controls as follows: 36.003 DISP 36.003 DISP ÿ SLOPE ................................ - (button in) 36.004 EVAL B TRIGGER LEVEL control? 37.001 HEAD {} 37.002 HEAD {CH2, A trigger mode P-P AUTO, slope positive} # This jmp FSC is needed for Cancel and Repeat options during test run. 37.003 JMP 38.001 37.004 EVAL Dummy 38.001 HEAD Internal A and B Triggering Test 38.002 DISP Move the Fluke 6061A RF OUTPUT via a 50ê feedthru 38.002 DISP termination to the UUT CH2 input connector. 38.003 DISP Set the UUT VERTICAL controls as follows: 38.003 DISP 38.003 DISP ÿ MODE ........................................... CH2 38.004 DISP Set the UUT HORIZONTAL controls as follows: 38.004 DISP 38.004 DISP ÿ MODE ............................................. A 38.005 DISP Set the UUT B TRIGGER controls as follows: 38.005 DISP 38.005 DISP ÿ SLOPE ............................... + (button out) 38.006 6060 100MH 100mVpp -D S 38.007 MESS Can a stable display be obtained by adjusting the 38.008 EVAL A TRIGGER LEVEL control? 39.001 HEAD {} 39.002 HEAD {CH2, A trigger mode P-P AUTO, slope negative} # This jmp FSC is needed for Cancel and Repeat options during test run. 39.003 JMP 40.001 39.004 EVAL Dummy 40.001 HEAD Internal A and B Triggering Test 40.002 6060 100MH 100mVpp -D S 40.003 DISP Set the UUT A TRIGGER controls as follows: 40.003 DISP 40.003 DISP ÿ SLOPE ................................ - (button in) 40.004 EVAL A TRIGGER LEVEL control? 41.001 HEAD {} 41.002 HEAD {CH2, A trigger mode NORM, slope negative} # This jmp FSC is needed for Cancel and Repeat options during test run. 41.003 JMP 42.001 41.004 EVAL Dummy 42.001 HEAD Internal A and B Triggering Test 42.002 6060 100MH 100mVpp -D S 42.003 DISP Set the UUT A TRIGGER controls as follows: 42.003 DISP 42.003 DISP ÿ A TRIGGER MODE ................................ NORM 42.004 EVAL A TRIGGER LEVEL control? 43.001 HEAD {} 43.002 HEAD {CH2, A trigger mode NORM, slope positive} # This jmp FSC is needed for Cancel and Repeat options during test run. 43.003 JMP 44.001 43.004 EVAL Dummy 44.001 HEAD Internal A and B Triggering Test 44.002 6060 100MH 100mVpp -D S 44.003 DISP Set the UUT A TRIGGER controls as follows: 44.003 DISP 44.003 DISP ÿ SLOPE ............................... + (button out) 44.004 EVAL A TRIGGER LEVEL control? 45.001 HEAD {} 45.002 HEAD {CH2, B trigger mode NORM, slope positive} # This jmp FSC is needed for Cancel and Repeat options during test run. 45.003 JMP 46.001 45.004 EVAL Dummy 46.001 HEAD Internal A and B Triggering Test 46.002 6060 100MH 100mVpp -D S 46.003 DISP Set the UUT HORIZONTAL controls as follows: 46.003 DISP 46.003 DISP ÿ MODE ............................................. B 46.004 DISP Set the UUT A TRIGGER controls as follows: 46.004 DISP 46.004 DISP ÿ A TRIGGER MODE ............................ P-P AUTO 46.005 EVAL B TRIGGER LEVEL control? 47.001 HEAD {} 47.002 HEAD {CH2, B trigger mode NORM, slope negative} # This jmp FSC is needed for Cancel and Repeat options during test run. 47.003 JMP 48.001 47.004 EVAL Dummy 48.001 HEAD Internal A and B Triggering Test 48.002 6060 100MH 100mVpp -D S 48.003 DISP Set the UUT B TRIGGER controls as follows: 48.003 DISP 48.003 DISP ÿ SLOPE ................................ - (button in) 48.004 EVAL B TRIGGER LEVEL control? 49.001 MESS 49.002 HEAD 49.003 DISP [27][91]1mDisconnect the test set-up. 49.004 END